
I-E67-TB82FF-2A February 13, 2002
6
Fault Code Fault Text
String
Corrective Action
PC12 TEMP.AD See PC8 corrective actions. If all items check out properly, replace Input PCB
Assembly. Transmitter can be used; however, the Temperature Compensation must be
set to Manual and a Temperature Calibration should be conducted in order to set the
Manual Temperature Value close to the operating conditions.
PC20 BAD.SEE
PC21 NO.F.CAL
PC30 R0.F.CAL
PC31 R1.F.CAL
PC32 R2.F.CAL
PC33 R3.F.CAL
PC34 R4.F.CAL
PC35 G0.F.CAL
PC36 G1.F.CAL
PC37 G2.F.CAL
PC38 G3.F.CAL
PC39 G4.F.CAL
PC40 W0.F.CAL
PC41 W1.F.CAL
PC42 W2.F.CAL
PC43 W3.F.CAL
PC44 W4.F.CAL
PC45 BA.F.CAL
PC46 PT.F.CAL
PC47 RT.F.CAL
PC50 R0.CHKS
PC51 R1. CHKS
PC52 R2. CHKS
PC53 R3. CHKS
PC54 R4. CHKS
PC55 G0. CHKS
PC56 G1. CHKS
PC57 G2. CHKS
PC58 G3. CHKS
PC59 G4. CHKS
PC60 W0. CHKS
PC61 W1. CHKS
PC62 W2. CHKS
PC63 W3. CHKS
PC64 W4. CHKS
PC65 BA.CHKS
PC66 PT.CHKS
PC67 RT.CHKS
PC70 GL.AD
PC72 DS.AD
Contact Factory.
Calibration Diagnostic Messages
The transmitter performs automatic efficiency and
offset calculations relative to a theoretically perfect
conductivity and/or temperature sensor during each
calibration cycle. Calibration history is retained for
future interrogation using the Edit Calibrate State.
The calibration constants that are displayed are Slope
and Offset for the Process Variable and Temperature.
A Slope of less than 0.2 or greater than 5 indicates a
potentially bad process calibration point or poorly
performing sensor. In these cases, the text string
BAD.CAL (bad calibration) is displayed in the
secondary display region. The user is returned to the
beginning of the calibration cycle after the bad
calibration has been reported.
An Offset value of less than -20 µS/cm for Sensor
Group A, -4 µS/cm for Sensor Group B, and -0.800
µS/cm or greater than 20 µS/cm for Sensor Group A,
4 µS/cm for Sensor Group B, and 0.800 µS/cm for
Commentaires sur ces manuels